Walking Boot Design: A Gait Analysis Study
|
| Fabian E. Pollo, PhD
Tracy L. Gowling
Robert W. Jackson, MD |
| From
the Department of Orthopedic Surgery, Baylor University Medical Center, Dallas, TX
This work was
supported by a grant from Bledsoe Brace Systems Inc.
Reprint requests:
Fabian Pollo, PhD, Dept of Orthopedic Surgery, Baylor University Medical Ctr, Dallas, TX
75246 |
|
| ABSTRACT |
| This
study investigated the effect of various short-leg walking boot designs on walking. Four
commercially available walking boots and a synthetic walking cast were tested using
three-dimensional gait analysis to determine their effect on a normal individuals
gait pattern. Temporal-spatial parameters and lower limb kinematics and kinetics were
analyzed for 10 normal subjects. The synthetic walking cast was the most different when
compared with the shoe condition with respect to all the parameters tested. The Bledsoe
walking boot was the only boot that was not significantly different from the shoe
condition for all parameters tested. Short lower-leg walkers have
been designed as an alternative to traditional cast immobilization. The indications for
use of lower-leg walkers include severe sprains, ankle and foot fractures, ad
post-surgical stabilization after tendon or ligament repairs. These orthopedic walkers are
light weight, convenient to use, and removable to perform range of motion exercises or
inspect the injured extremity. Short-leg walkers are as effective as walking casts in
healing ankle fractures, and patients treated with short-leg walkers experience
significantly less edema, tenderness, and joint stiffness after 6 weeks of immobilization.1
In recent years, the
cost of short-leg walkers has decreased substantially and even reached levels below the
cost of synthetic or traditional plaster walking casts. The degree of immobilization is
most likely more stringent in a cast; however, unless this is a requirement, the numerous
advantages short-leg walkers have over casting make them appealing.
To date, no studies
have investigated the effect short-leg walkers have on an individuals gait pattern.
Previous studies have investigated only foot pressures in individuals wearing these
orthoses.2,3 Because these orthoses sometimes are worn for lengthy periods (up
to 6 months), a normal gait pattern while wearing these walkers should be an important
consideration.
This study compared
the gait pattern in normal individuals wearing four commercially available short-leg
walkers and a synthetic walking cast to determine which orthosis allowed the individual to
simulate the most normal gait pattern. |
| MATERIALS AND
METHODS |
| Patient
Population. Six men and four women, with a mean age of 27.2±3.6 years, mean height
of 171.9±10.4 cm, and a mean weight of 74.7±13.9 kg, were recruited for this study.
Subjects were included only if they had no prior history of lower extremity injuries or
surgeries. All subjects gave informed consent prior to participation in the study. Subjects wore shorts and a
T-shirt to allow for maximum exposure of their legs and pelvic area during testing. Only
one side was tested for each subject. The side to be tested was randomly chosen just prior
to testing.
Gait Measurements.
Kinematic parameters were collected using a 5-camera Motion Analysis System (Motion
Analysis Corporation Inc, Santa Rosa, Calif) that recorded at 60 Hz. Ground reaction
forces were collected with an AMTI OR6-5 force platform (Advanced Medical Technology, Inc,
Newton, Mass) embedded in the center of an 8-m walkway and recorded at 1000 Hz.
Inverse dynamics were
used to derive kinetic information from the kinematic and ground reaction force data.
Reflective markers were secured to anatomical locations with adhesive tape using the Helen
Hayes Marker configuration.4 Markers were placed bilaterally over the
anterosuperior iliac spine, lateral knee epicondyle, lateral malleolus, directly on the
shoe or walking boot over the head of the third metatarsal, and the posterior portion of
the heel at the same level as the marker over the third metatarsal head.
Short 9.5-cm and
6.5-cm wands, with markers attached to the ends, were secured laterally to the mid-thigh
and mid-shank regions, respectively, in line with the greater trochanter and lateral knee
epicondyle marker for the thigh wand, and in line with the lateral knee epicondyle and
lateral malleolus markers for the shank wand. A single marker was placed over the space
between L5 ans S1 to identify the movement of the pelvis, in conjunction with the two
anterosuperior iliac spine markers.
Each marker location
was identified carefully with a surgical pen after attachment, to keep the locations of
the markers fixed between the various conditions to eliminate errors with marker
placement. Subjects were instructed to wear soft-soled shoes with a medium heel height on
the day of testing. The mean heel height for the group test was 1.91±0.29 cm. The first
condition tested was with the subjects own shoes and was considered the
normal baseline pattern against which the remaining conditions were compared.
Description of
the Short-Leg Walkers. Four commercially available short-leg walkers and a synthetic
short-leg walking cast were tested. The short-leg walkers were the:
- Bledsoe Walking Boot
(Bledsoe Brace System, Inc, Grand Prairie, Tx),
- Three-D Orthopedic
Samson (DeRoyal, Powell, Tn),
- Royce Equalizer (Royce
Medical/Center Orthopaedics, Camarillo, Ca), and
- Cam Walker (Zinco
Industries Inc, Pasadena, Ca).
The Bledsoe Walking
Boot retails for about $50 and is constructed of lightweight aluminum with side arms. The
boot has a heel height of approximately 1.9 cm, with about 1.5 cm of that comprising a
soft rubber sole. The sole of this boot has a rocker-shaped bottom to approximate the
three rocker motions of the ankle during gait. The sole is narrower at the heel (5.4 cm)
and progressively expands toward the toes (8.9 cm). According to the manufacturers
literature, The Bledsoe Walking Boot is designed as a cast replacement for bracing
fractures from the middle of the tibia or fibula to the mid-foot, and for second- and
third- degree ankle sprains.
The Three-D
Orthopedic Samson retails for approximately $45 and consists of an injected, molded
plastic shell with fixed plastic arms as part of the midsole. This boot has wide
medial/lateral paddles toward the upper calf portion. Heel height is approximately 3.8 cm,
with about 6 mm of that comprising a hard rubber sole. The Samson boot has a relatively
flat bottom sole with a figure-eight shape. The heel is approximately 8.6 cm wide, the
middle section is approximately 7 cm wide, and the toe is approximately 8.9 cm wide.
According to the manufacturers literature, the Three-D Samson Boot is indicated for
stable ankle and foot fractures, severe ankle sprains, and postoperative open reduction
internal fixation.
The Royce Equalizer
retails for about $55 and consists of injected, molded, glass-filled plastic, with plastic
arms bonded to the shell. This boot has a wide foot plate (12.7 cm) that is nearly uniform
throughout the entire sole. The heel height of this boot is approximately 4.2 cm, with a
3-mm hard rubber sole and a curved bottom sole. According to the manufacturers
literature, the Royce Equalizer Walker is indicated for soft-tissue injuries (grade 2 and
3 sprains), stable fractures, and postoperative use.
The Cam Walker
retails for approximately $72 and consists of an injected, molded plastic shell with twin
steel uprights and two hex screw adjusters for range of motion. This boot has a 5.4-cm
heel height with a 3-mm hard rubber sole. The Cam Walker boot has a curved bottom sole
with a figure-eight shape. The heel is approximately 7.9 cm wide, the middle is
approximately 7.6 cm wide, and the toe is approximately 9.2 cm wide. The
manufactures literature for Cam Walker does not include indications for the boot.
Small, medium, and
large sizes for each walker were available to accommodate all subject sizes. The synthetic
walking cast used was Delta-Lite Casting Tape (Johnson & Johnson Orthopedics) and was
applied by the same certified orthotist. A simple cast shoe was used in conjunction with
the synthetic cast during testing.
Subject testing.
The testing sequence for the walkers and cast was randomized. At the start of each
condition, the randomly selected walker or cast was placed onto the subjects
randomly selected limb by the same experienced kinesiologist or orthotist. Care was taken
not to disturb the markers between conditions. Only the markers on the tested leg below
the knee needed to be moved from condition-to-condition.
After application of
the walker or cast, the lateral malleolus marker was replaced as close as possible over
the location of the ankle malleolus. The shank wand was placed over the walker or cast, in
line with the lateral knee epicondyle and lateral malleolus markers. The markers on the
foot were placed as close as possible over the anatomical locations, but the foot markers
were not critical since the walkers and cast constrained the ankle in neutral; therefore,
the kinematic variables about the dorsiflexion/plantarflexion axis were not subsequently
important.
After the walker or
cast and markers were placed on the subject, the subject walked for several minutes to get
accustomed to that particular orthosis. Once the subject was ready to proceed with the
orthosis, five trials for that condition were recorded for averaging and further
statistical analysis. Subjects were allowed to select their own walking speed and were
allowed to rest when they felt the need.
Temporal-spatial
parameters included calculations of velocity, cadence, and stride length. The kinematic
parameters measured for the pelvis included obliquity, anteroposterior tilt, and
internal/external rotation. For the hip and knee, flexion/extension and
abduction/adduction kinematics were measured. Foot progression angle was the only
parameter measured at the ankle, since the walking boots and synthetic cast locked the
ankle at neutral in the dorsiflexion/plantarflexion and inversion/eversion planes.
External moments were computed for the hip and knee about the flexion/extension and
abduction/adduction planes.
Statistical
Analysis. Difference curves were computed for all kinematic and kinetic parameters.
Difference curves were calculated by subtracting the mean of the baseline shoe trial from
the mean of each boot and cast condition for each subject. Therefore, the difference
curves for each parameter for all 10 subjects represented a normalized effect of that
particular walking boot or cast with respect to the baseline shoe trial.
A statistical
technique called boot-strapping5 was then used to computer the mean ans 95%
confidence limits of the difference curves for the group for each parameter. When the
upper or lower 95% confidence limit for a particular parameter crossed the zero line
during the gait cycle, that portion was considered statistically significant.
Temporal-spatial parameters were analyzed using the Mann-Whitney test at alpha=.05. |
| RESULTS |
| Temporal-Spatial
Parameters. Temporal-spatial parameters are summarized in the Table. The synthetic
cast was the only orthosis that demonstrated a significant reduction in velocity and
cadence compared with the baseline shoe trials (P<.05). Stride length was not
significantly different for all tested conditions and the baseline shoe trials. Pelvic Kinematics.
Pelvic kinematics during gait were virtually unchanged with the synthetic walking cast.
Mean angular differences between the cast trials and the baseline shoe trials ranged from
1.3 to 1.2 for obliquity, 3.3 to 0.9 for anterior tilt, and 1.3 to 2.8°
for internal rotation.
Pelvic kinematics
also were close to baseline for all walking boots tested, with the exception of the
Three-D Orthopedic Samson boot, which demonstrated a slight increased anterior pelvic tilt
between 53% and 56% of the gait cycle, just before swing phase. Overall, the mean angular
differences between the walking boots and the baseline shoe trials ranged from 1.5 to
1.2 for obliquity, 2.6 to 0.4 for anterior tilt, and 3.1 to 2.2 for
internal rotation.
Hip Kinematics.
Hip kinematics were not significantly different for the cast or any of the walking boots
compared with the baseline shoe trials ranged from 1.9 to 2.6 for abduction and 5 to
1.8 for flexion.
Knee Kinematics.
Knee flexion was significantly reduced during the cast trials compared with the baseline
shoe trials for most of the stance phase (13% to 52% of gait cycle). Knee
abduction/adduction kinematics were unchanged for the cast trials, and neither flexion nor
adduction kinematics were altered by any walking boot. Average angular differences for the
four walking boots compared with the baseline shoe trials ranged from 1.2 to 3.2 for
adduction and 6.2 to 8.4 for flexion.
Foot Progression
Angle. Foot progression angle was not significantly different in the cast or any
walking boot compared with the baseline shoe trials. Average angular differences for the
cast and four walking boots compared with the baseline shore trials ranged from 11.1 to
7.1 for internal rotation. All five orthosis demonstrated decreased external
rotation during stance phase with an increased external rotation just after toe off into
swing phase; however, no changes were statistically significant.
Hip Moments.
Flexion/extension moments about the hip were not significantly different for the cast or
any of the walking boots compared with the baseline shoe trials. However, the cast ana all
of the walking boots, except the Bledsoe Brace, had statistically significant differences
in the hip abduction moment compared to the shoe trials. With the cast, the external
abduction moment was reduced during 15% to 19.2% of the gait cycle. The Three-D
Orthopedic, Cam Walker, and Royce Equalizer boots demonstrated decreases in the external
hip abduction moment during 38.4% to 41.4%, 12.6% to 16.8%, and 32.4% to 39.6% of the gait
cycle, respectively.
Knee Moments.
External knee adduction moments were significantly changed for the cast, the Three-D
Orthopedics Samson, and the Cam Walker from 10.8% to 19.8%, 32.4% to 45.6% and 6% to 13.2%
of the gait cycle, respectively. The typical knee flexion moment during stance phase was
significantly reduced while in the cast from 9% to 33.6% of the gait cycle. The Cam Walker
and Royce Equalizer also demonstrated some small but significant increases inn the
external flexion moment during stance compared to the baseline shoe trials. The Cam Walker
was significant between 25.8% and 35.4% of the gait cycle, and the Royce Equalizer was
significant between 23.4% and 32.4% of the gait cycle. |
| DISCUSSION |
| The
synthetic walking cast was the only orthosis that significantly altered gait from the
baseline shoe trials. All of the walking boots performed better than the cast in allowing
subjects to more closely simulate their baseline shoe gait patterns. The Bledsoe Boot
performed the best, demonstrating no significant differences in any of the kinematic or
kinetic parameters with respect to the baseline shoe trials. Knee kinematics differences
for all of the walking boots were insignificant for both the flexion/extension and
abduction/adduction planes. Similarly, foot progression angle was not altered in either
the walking cast or any of the walking boots during the gait cycle. The trend of the foot
progression angle for all orthoses was a slight decrease in external rotation during
stance and an increase in external rotation during swing.
The external moments about the hip also were not significantly different for any of the
orthoses in the flexion/extension plane. However, in the abduction/adduction plane, the
cast abs all of the walking boots except the Bledsoe boot had significant differences
compared with the baseline shoe trials. These differences occurred during the higher
loading portions of the stance phase. This alteration could lead to some abnormal joint
loading that eventually may adversely affect the articular surfaces at the hip.
These differences probably can be attributed to the sole design of the boots.
Structurally, all of the boots are basically the same, with a padded foot shell and
semi-rigid side arms running up the shank. The biggest difference between the boots is the
sole design and thickness. Since the kinetics at the hip are calculated after the knee
joint kinetics, it was not surprising to see these differences when the external adduction
moments about the knee were significantly affected in two of the four boots.
The external adduction moment about the knee was significantly reduced for the
synthetic cast, the Three-D Orthopedic Samson, and the Cam Walker during a small portion
of the stance phase. The external adduction moment at the knee is crucial for maintaining
appropriate loads on the medial and lateral compartments of the knee, and is partly a
result of the knee adduction/abduction angulation and the ground reaction component in the
frontal plane.
These differences were not attributed to the knee adduction/abduction angulation, as
that parameter was nearly identical for all boots. Therefore, the changes must arise from
the ground reaction forces, which the authors believe is a direct result of the
differences in the sole design. The Three-D Orthopedic and Cam Walker both have a
figure-eight design with a thin hard rubber sole. The rigidity of the sole in these two
boots limits the amount of mediolateral deviation from the inherent contour of the boots.
The Royce Equalizer has a wide sole plate possibly controlling mediolateral play, and the
Bledsoe Walking Boot with its soft rubber sole, allows for a natural positioning of the
foot during stance.
For external flexion/extension moments, apart from the synthetic cast, the Cam Walker
and Royce Equalizer demonstrated increased flexion moments about the knee during the
muscle portion of the stance phase. This increase in the external flexion moment would
require greater effort from the knee extensors and probably lead to higher than normal
forces at the tibiofemoral and patellofemoral joints, and also slightly increase the
physical demand of ambulation. Again, these differences probably can attribute to sole
design.
The Cam Walker and Royce Equalizer both have a curved sole design, but with a thin,
hard rubber sole. The sole design during mid-stance may not position the lower limb in the
same manner as the natural ankle. This tends to increase the knee flexion angle and
possibly increase the posterior ground reaction force, which both contribute to the
increase in the external knee flexion moment. The Three-D Orthopedic Samson boot has a
relatively flat sole, thereby allowing the lower shank to stay perpendicular during
mid-stance. The Bledsoe Walking Boot has a three rocker sole design and a soft rubber sole
that allows the shank to orient itself more naturally during stance. In addition, of the
four walking boots, the Bledsoe Walking Boot has the lowest profile heel that most closely
matches the heel height of a walking shoe, which also could be a factor in allowing the
wearer to simulate a more normal gait pattern in that particular boot. |
| CONCLUSION |
| Walking
boots offer several advantages over a synthetic walking cast. First, walking boots can be
removed easily to allow range-of-motion exercises and direct visualization of the injured
area. Second, the cost of these walking boots have reached levels well below their
synthetic cast counterpart. Third, walking boots allow both the injured limb and the boot
to be cleaned when necessary. Finally, as demonstrated in this study, walking boots have a
less adverse effect on kinematic and kinetic gait patterns than a synthetic walking cast. |
| EDITORIAL
DISCUSSION |
| ORTHOPEDICS:
Normal subjects were tested in this study. However, the boots are applied only to
patients. How do data from normal subjects carry over to patients, who will not have
normal gait patterns to begin with? Pollo et al: These boots are
intended for patients who have foot and ankle problems. The purpose of a walking boot is
to keep the foot and ankle in a relatively neutral position during ambulation with good
support to distribute the load over a greater area. The gait parameters measured in this
study were focused more at proximal joints (ie the knee, hip, and pelvis) and the effects
that walking boots have on those areas. Therefore, these data do carry over to patients
since ankle motion is eliminated in the boots and compensations must then be made at the
proximal joints to ambulate.
Pain is the only factor that may become an issue, since if there is some pain or
discomfort present while walking in these boots, then other compensations may be made by
the patient to reduce the load and subsequent pain on the foot/ankle. This issue is beyond
the scope of this study as we did not investigate the pressures or loads on the foot/ankle
with each boot to determine which boot minimizes these factors during walking. The authors
assumed that patients who use these boots will have normal knee and hip function without
any significant pain while walking.
The boots serve the purpose of immobilizing the limb while making skin care and
rehabilitation easier. Because these boots can be worn for long periods of time, their
effects on gait patterns at other joints can become an important issue.
ORTHOPEDICS: Would the effects sometimes be more favorable and other
times less favorable depending on the disability, surgery, and treatment of each patient
category?
Pollo et al: The effects should always be favorable for any foot or
ankle condition for which a walking boot is prescribed. However, we have not investigated
the effect of pain in these boots and how that affects ambulation in these various boots.
If there is severe pain while walking in the se boots, then it should be managed by
analgesics or with nonweight bearing, and whether ambulation is normal in these boots is
probably not of concern. Therefore, these data show that any foot or ankle condition that
is not accompanied by pain would benefit from a walking boot that allowed a more normal
gait pattern. Other factors that also could come into play but were not addressed in this
study include other neurologic or musculoskeletal problems that could cause gait
adaptations that may or may not benefit from a walking boot that allowed a more normal
gait pattern in a healthy individual.
ORTHOPEDICS: The differences between the Bledsoe Boot and the other
walking boots were minor. Can you comment on the differences that made the Bledsoe Boot
perform slightly better than the other walking boots?
Pollo et al: The main difference between the Bledsoe Walking Boot and the other walking
boots tested was the sole, design. The Bledsoe Walking Boot was the only boot with a
three-part rocker design that mimics the natural ankle movement during the gait. In
addition, The Bledsoe Walking Boot uses a thick soft rubber sole, providing more
cushioning at heel strike and better transition through stance phase and toe-off, which is
typically achieved through the ankle joint but has been lost as a result of the fixed
ankle joint position.
Although these differences are minor, slight deviations from normal can lead to
abnormal joint loading and increased physical demand. Therefore, is an individual has pain
ata the knee or hip, or is in poor physical condition, one would prefer not to increase
the loads across those painful joints or increase ones physical demand for walking. |
TABLE
Temporal-spatial parameters |
| Orthosis |
Velocity(cm/sec) |
Cadence(steps/min) |
Stride
Length(cm) |
| Shoe |
135.6 ±11.3 |
111.4 ±8.2 |
146.1 ±9.9 |
| Synthetic cast |
119.2 ±16.5 |
101.2 ±7.7* |
141.9 ±11.6 |
| Bledsoe |
129.5 ±16.5* |
107.4 ±7.1 |
146.0 ±7.9 |
| Three-D Samson |
132.1 ±14.8 |
107.1 ±8.1 |
147.9 ±8.8 |
| Cam Walker |
133.2 ±16.0 |
107.4 ±7.8 |
148.6 ±8.2 |
| Royce Equalizer |
131.0 ±14.9 |
106.1 ±7.3 |
149.1 ±8.4 |
| *Significant
difference compared with shoe condition, P<.05. |
| REFERENCES |
| 1.
Polakoff DR, Pearce SM, Grogan DP, Burkhead WZ. The orthotic treatment of stable ankle
fractures. Orthopedics. 1984; 1712-1715.
2. Birke JA, Nawoczenski DA. Orthopedic walkers: effect on plantar pressures. Clinical
Prosthetics and Orthotics. 1988; 12:74-80.
3. Lavery LA, Vela SA, Lavery DC, Quebedeaux TL. Reducing dynamic foot pressures in
high risk diabetic subjects with foot ulcerations. A comparison of treatments. Diabetics
Care. 1996; 19:818-821.
4. Kadaba MP, Ramakrishnan HK, Wootten ME. Measurement of lower extremity kinematics
during level walking. J Orthop Res. 1990; 8:383-392.
5. Olshen R, Biden E, Wyatt M, Sutherland D. Gait analysis and the bootstrap. Annals of
Statistics. 1989; 17:1419-1440. |
|